CALIBRATION OF VOLUMETRIC GLASSWARE


EXPERIMENT # 2

ADORNA JR., JOEMER A. PITAGAN, PAULA JESSIKA C.

MALAYAN COLLEGES LAGUNA

EXPERIMENT # 2 CALIBRATION OF VOLUMETRIC GLASSWARES

I.

OBJECTIVES To identify different types of volumetric glassware; To discuss the importance of calibrating volumetric instruments; To calibrate volumetric glassware; and To use volumetric glassware properly.

II.

LABORATORY EQUIPMENT / INSTRUMENTS / REAGENTS Equipment/Accessories 25 mL volumetric pipet 10 mL volumetric pipet 50 mL buret Aspirator Analytical balance Iron stand 50 mL beaker 250 mL beaker 125 mL Erlenmeyer flasks Thermometer Parafilm Buret holder Distilled water in wash bottle Quantity 1 1 2 1 1 1 2 2 2 1 1 1 1

III.

DISCUSSION OF FUNDAMENTALS Introduction In the life of a simple average person, he measures everyday things with a little something called "estimate", especially to liquids, things, and the like. But to scientists, estimation is out of the question. Scientists, especially chemists, do not take second chances when measuring volatile reactive liquids into their reactions. A minute addition or subtraction to it might cause explosions, unexpected or non-visual reactions. This is very critical especially in the pharmaceutical or in the Experiment 2: Calibration of Volumetric Glassware 1|Page

MALAYAN COLLEGES LAGUNA

EXPERIMENT # 2 CALIBRATION OF VOLUMETRIC GLASSWARES

food industry. If you accidentally killed a person with a confident estimate, then chances are you are on the verge of your pulling conscience, if you have one. Like the first experiment, measuring volumes are as fundamental as measuring mass. But they are a little more crucial when it comes to measuring liquid volume. Unlike solids, they more significantly respond to changes in pressure and temperature, which is particularly dynamic, even in a laboratory setting. In doing so, volumetric devices for measurements are used. The type of volumetric device to be used for a particular measurement considers four factors: general goal of the volume measurement, volume or range of volumes to be measured, degree of reliability needed for the measurement and number of measurements to be made. Calibration of these devices is important, as the reasons are stated above. Theory Volume like mass is another fundamental property of matter that is commonly determined in analytical measurements. For solids, volume can be obtained through calculations of the objects dimensions. For liquid materials, volume can be determined by determining the volume of the container the liquid occupies. Most common laboratory glassware like beakers, Erlenmeyer flasks, and test tubes serve as containers for mixing, handling, and heating solutions but are not designed for accurate volume determinations. Volumetric devices used for analytical measurements include volumetric flask, volumetric pipets or transfer pipet, burets, micropipets, and syringes (Hage and Carr 2011). Calibration of volumetric devices is very important especially when the device is recently acquired or when the device will be used at a temperature different from the temperature it was initially calibrated. This is because glassware will contract or expand with a change in temperature. In addition, water expands about 0.02% per degree around 20C (Christian 2004). Therefore, the true volume is different from the volume that is read from the container. The true volume can be achieved by calculation considering buoyancy effects and measuring the mass of water that is contained by the volumetric device, and then calculating the volume of water that was present using the known density of water at that temperature (Hage and Carr 2011). Application Practically this experiment must be done first, among all others. This is practically because it main application is to ensure the accuracy of the volumetric glassware, that will be frequently used Experiment 2: Calibration of Volumetric Glassware 2|Page

MALAYAN COLLEGES LAGUNA

EXPERIMENT # 2 CALIBRATION OF VOLUMETRIC GLASSWARES

in the proceeding experiments. If this experiment gives out bad values, even though random and systematically errors have been properly avoided and accounted for, chances are that the remaining experiments that will follow onto this will definitely carry on this same type of error, an error that is within the apparatus, not on the user itself.

IV.

METHODOLOGY
Glasswares were cleaned with detergent and lots of tap water. They were then rinsed with distilled water.

200-mL distilled water was placed in a 250-mL beaker and was allowed to equilibrate in room temperature.

Distilled water bottle was allowed to equilibrate to room temperature. It was used for the calibration of the burette.
Figure 2.1. Glassware and sample preparation

Experiment 2: Calibration of Volumetric Glassware

3|Page

MALAYAN COLLEGES LAGUNA

EXPERIMENT # 2 CALIBRATION OF VOLUMETRIC GLASSWARES

A clean and dry Erlenmeyer flask with a stopper was weighed in an analytical balance. The weight was recorded.

Actual/ true volume was taken using: Mass of H2O with buoyancy effect = mass of H2O * buoyancy correction (eq. 1)

Burette was filled with distilled water up to the 0.00-mL mark. It was ensured that no leakages or bubbles were found.

True volume = (corrected mass of H2O)/(density of H2O at specified temperature). (eq. 2)

Initial volume was read to the nearest 0.01 using a meniscus reader.

Correction value for each apparent volume was taken using: Correction value = true volume apparent volume (eq. 3)

About 10-mL of water was transferred from the burette to the preweighed Erlenmeyer flask. Delivered water = apparent volume.

A stopper was put on the flask with water. Mass was recorded. Step was repeated until water in buret was discarded until 50-mL mark carefully.

Entire procedure was repeated with 20 aliquots of water per delivery. Same was done for 30-mL, 40-mL, and 50-mL aliquots of water.

Values were plotted: correction values on y-axis and apparent volume on x-axis. Differences in mass between two consecutive mass weighings were taken as water delivered.
Figure 2.2. Calibration of 50-mL buret

Experiment 2: Calibration of Volumetric Glassware

4|Page

MALAYAN COLLEGES LAGUNA

EXPERIMENT # 2 CALIBRATION OF VOLUMETRIC GLASSWARES

For 25-mL pipet: A clean and dry 125-mL Erlenmeyer flask with stopper was weighed on an analytical balance. Recorded as: initial mass. For 10-mL pipet: A clean, dry 50mL beaker with parafilm was weighed on an analytical balance. Recorded as: initial mass.

Difference in mass between each set of two consecutive mass measurements were obtained to determine mass of water delivered in each trial.

Get true volume using eq. 1 25-mL/10-mL water was transferred from the pipet, allowing water to run out, with the tip of the pipet touching the side of the beaker. It was allowed 7-10 seconds to drain. The mass of the beaker and added water was recorded. Recorded: water delivered = apparent volume.

Do entire procedure using 10-mL pipet.

This addition was repeated three more times without discarding water in beaker or flask. Recorded for every transfer: apparent volume and mass of container plus added water.
Figure 2.3. Calibration of 25-mL and 10-mL pipets

The following were calculated for calibrated pipets: a. the mean volume, b. standard deviation, c. relative standard deviation (%RSD), d. 95% confidence interval, e. % relative error (theoretical volume assumed was exactly 25-mL and 10-mL for pipets)

Experiment 2: Calibration of Volumetric Glassware

5|Page

MALAYAN COLLEGES LAGUNA

EXPERIMENT # 2 CALIBRATION OF VOLUMETRIC GLASSWARES

V.

DESCRIPTION OF THE APPARATUS / SET - UP

Figure 2.4. Volumetric flask.

Figure 2.5. Volumetric pipet

Figure 2.6. Burette (buret)

A volumetric flask is an instrument that is used to contain an accurate amount of liquid. A typical volumetric flask that is used in our laboratory measures 500 mL of liquid, which has a Experiment 2: Calibration of Volumetric Glassware 6|Page

MALAYAN COLLEGES LAGUNA

EXPERIMENT # 2 CALIBRATION OF VOLUMETRIC GLASSWARES

tolerance of 0.2 mL. This tolerance equates to a relative uncertainty of about 400 ppm (parts per million). A volumetric pipet is an instrument that is used to transfer accurate amounts of liquid. There are many types of volumetric pipets, among which are used in this laboratory where the 10-mL and the 25-mL ones. The volumetric pipet has only one graduation, and in doing so it can only deliver one accurate measure at a time. A burette is a laboratory apparatus that is mainly used for quantitative chemical analyses of liquids. It consists of a long, graduated glass tube with a stopcock (in a liquid burettes case, on the bottom) that is handled by a burette clamp, which is connected to an iron stand. The volume that the burette dispenses is controlled by the stopcock, and is accurately measured by the graduations of the glass tube.

VI. I.

DATA SHEET GLASSWARE AND SAMPLE PREPARATION

Table 1. Water temperature Container 250 mL beaker Distilled water bottle Temperature (C) 24C 24C

Experiment 2: Calibration of Volumetric Glassware

7|Page

MALAYAN COLLEGES LAGUNA

EXPERIMENT # 2 CALIBRATION OF VOLUMETRIC GLASSWARES

II. CALIBRATION OF 50 mL BURET Table 2. Calibration of 50 mL buret Conditions Mass of Erlenmeyer flask, g 10-mL delivery (1st) Final volume, mL Initial volume, mL Volume used, mL Mass of flask + 10 mL water, g Mass of water, g 10-mL delivery (2nd) Final volume, mL Initial volume, mL Volume used, mL Mass of flask + 10 mL water, g Mass of water, g 10-mL delivery (3rd) Final volume, mL Initial volume, mL Volume used, mL Mass of flask + 10 mL water, g 30.0 20.0 10.0 143.28 30.0 20.0 10.0 147.54 20.0 9.9 10.1 133.2 10.05 20.0 10.0 10.0 137.59 10.02 9.9 0.00 9.9 123.19 9.41 10.0 0.00 10.0 127.57 10.05 Trial 1 113.79 Trial 2 117.52

Experiment 2: Calibration of Volumetric Glassware

8|Page

MALAYAN COLLEGES LAGUNA

EXPERIMENT # 2 CALIBRATION OF VOLUMETRIC GLASSWARES

Mass of water, g 10-mL delivery (4th) Final volume, mL Initial volume, mL Volume used, mL Mass of flask + 10 mL water, g Mass of water, g 10-mL delivery (5th) Final volume, mL Initial volume, mL Volume used, mL Mass of flask + 10 mL water, g Mass of water, g Mass of Erlenmeyer flask, g 20-mL delivery (1st) Final volume, mL Initial volume, mL Volume used, mL Mass of flask + 20 mL water, g Mass of water, g 20-mL delivery (2nd)

9.96

9.95

40.0 30,0 10.0 153.28 10.02

40.1 30.0 10.1 157.61 10.07

50.0 40.0 10.0 163.27 10.01 113.79

50.0 40.1 9.9 167.51 9.9 117.52

19.9 0.00 19.9 137.37 19.85

20.0 0.00 20.0 137.36 19.84

Experiment 2: Calibration of Volumetric Glassware

9|Page

MALAYAN COLLEGES LAGUNA

EXPERIMENT # 2 CALIBRATION OF VOLUMETRIC GLASSWARES

Final volume, mL Initial volume, mL Volume used, mL Mass of flask + 20 mL water, g Mass of water, g Mass of Erlenmeyer flask, g 30-mL delivery (1st) Final volume, mL Initial volume, mL Volume used, mL Mass of flask + 30 mL water, g Mass of water, g Mass of Erlenmeyer flask, g 40-mL delivery (1st) Final volume, mL Initial volume, mL Volume used, mL Mass of flask + 40 mL water, g Mass of water, g Mass of Erlenmeyer flask, g 50-mL delivery (1st)

40.0 19.9 20.1 157.47 20.1 117.52

40.2 20.0 20.2 157.58 20.22 117.52

30.0 0.00 30.0 147.62 30.1 117.52

30.1 0.00 30.1 147.59 30.07 117.52

40.1 0.00 40.1 157.58 40.06 117.52

40.0 0.00 40.0 157.51 39.99 117.52

Experiment 2: Calibration of Volumetric Glassware

10 | P a g e

MALAYAN COLLEGES LAGUNA

EXPERIMENT # 2 CALIBRATION OF VOLUMETRIC GLASSWARES

Final volume, mL Initial volume, mL Volume used, mL Mass of flask + 50 mL water, g Mass of water, g

50.0 0.00 50.0 167.21 49.69

50.0 0.00 50.0 167.46 49.94

Table 3. Volumes for the 50 mL Buret Conditions 10-mL delivery (1st) Apparent volume , mL Mass of water, g Corrected mass, g True volume, mL Correction value, mL 10-mL delivery (2nd) Apparent volume , mL Mass of water, g Corrected mass, g True volume, mL Correction value, mL 10-mL delivery (3rd) 10.1 10.13 10.15 10,17 0.07 10 10.02 10.04 10.06 0.06 9.9 9.4 9.41 9.44 -0.46 10 10.05 10.07 10.10 0.10 Trial 1 Trial 2

Experiment 2: Calibration of Volumetric Glassware

11 | P a g e

MALAYAN COLLEGES LAGUNA

EXPERIMENT # 2 CALIBRATION OF VOLUMETRIC GLASSWARES

Apparent volume , mL Mass of water, g Corrected mass, g True volume, mL Correction value, mL 10-mL delivery (4th) Apparent volume , mL Mass of water, g Corrected mass, g True volume, mL Correction value, mL 10-mL delivery (5th) Apparent volume , mL Mass of water, g Corrected mass, g True volume, mL Correction value, mL 20-mL delivery (1st) Apparent volume , mL Mass of water, g Corrected mass, g

10 9.96 9.98 10.00 0

10 9.95 9.97 10.00 0

10 10 10.02 10.04 0.04

10.1 10.07 10.09 10.12 0.11

10 9.99 10.01 10.03 0.03

9.9 9.9 9.92 9.95 0.05

19.9 19.85 19.88

20.0 19.84 19.87

Experiment 2: Calibration of Volumetric Glassware

12 | P a g e

MALAYAN COLLEGES LAGUNA

EXPERIMENT # 2 CALIBRATION OF VOLUMETRIC GLASSWARES

True volume, mL Correction value, mL 20-mL delivery (2nd) Apparent volume , mL Mass of water, g Corrected mass, g True volume, mL Correction value, mL 30-mL delivery (1st) Apparent volume , mL Mass of water, g Corrected mass, g True volume, mL Correction value, mL 40-mL delivery (1st) Apparent volume , mL Mass of water, g Corrected mass, g True volume, mL Correction value, mL 50-mL delivery (1st)

19.93 0.03

19.92 -0.08

20.1 20.1 20.13 10.19 0.09

20.2 20.22 20.25 20.31 0.11

30.0 30.1 30.14 30.23 0.23

30.1 30.07 30.12 30.2 0.1

40.1 40.06 40.12 40.23 0.13

40.0 39.99 40.05 40.16 0.16

Experiment 2: Calibration of Volumetric Glassware

13 | P a g e

MALAYAN COLLEGES LAGUNA

EXPERIMENT # 2 CALIBRATION OF VOLUMETRIC GLASSWARES

Apparent volume , mL Mass of water, g Corrected mass, g True volume, mL Correction value, mL

50.0 49.69 49.77 49.9 -0.10

50.0 49.94 50.05 50.15 0.15

III.

CALIBRATION OF 25 mL AND 10 mL VOLUMETRIC PIPET

Table 4. Calibration of 25 mL volumetric pipet and 10 mL measuring pipet Conditions Mass of Erlenmeyer flask, g 25-mL delivery (1st trial) Volume delivered, mL Mass of flask + 25 mL water, g Mass of water, g Corrected mass, g True volume, mL 25-mL delivery (2nd trial) Volume delivered, mL 25.0 25.0 142.45 24.93 25.02 25.04 25-mL pipet 117.52 Conditions Mass of beaker, g 10-mL delivery (1st trial) Volume delivered, mL Mass of beaker + 10 mL water, g Mass of water, g Corrected mass, g True volume, mL 10-mL delivery (2nd trial) Volume delivered, mL 10.0 10.0 40.35 10-mL pipet 30.45

9.9 9.91 9.94

Experiment 2: Calibration of Volumetric Glassware

14 | P a g e

MALAYAN COLLEGES LAGUNA

EXPERIMENT # 2 CALIBRATION OF VOLUMETRIC GLASSWARES

Mass of flask + 25 mL water, g Mass of water, g Corrected mass, g True volume, mL 25-mL delivery (3rdtrial ) Volume delivered, mL Mass of flask + 25 mL water, g Mass of water, g Corrected mass, g True volume, mL

167.33 24.88 24.96 24.99

Mass of beaker + 10 mL water, g Mass of water, g Corrected mass, g True volume, mL 10-mL delivery (3rdtrial)

30.29

9.94 9.96 9.98

25.0 192.21 24.88 24.97 24.99

Volume delivered, mL Mass of beaker + 10 mL water, g Mass of water, g Corrected mass, g True volume, mL Statistical Analysis

10.0 60.28

9.99 10.01 10.03

Average volume Standard deviation RSD 95% confidence interval % relative error

25.01 0.029 0.12% 25.010.028 0.04%

Average volume Standard deviation RSD 95% confidence interval % relative error

9.96 0.05 0.54% 9.960.049

0.4%

Experiment 2: Calibration of Volumetric Glassware

15 | P a g e

MALAYAN COLLEGES LAGUNA

EXPERIMENT # 2 CALIBRATION OF VOLUMETRIC GLASSWARES

VII.

SAMPLE CALCULATIONS

For Table 3 ( )( )

VIII.

RESULTS AND DISCUSSIONS This experiment is all about calibration of laboratory glassware. Calibration is very much needed for the instruments, since it is one of the primary processes used in maintaining the instruments measuring accuracy. Calibration, hence, means the process of configuring an instrument (and accepting whatever environmental factors might be affecting it, even in a controlled laboratory setting) to provide certain results from a sample within an acceptable range (which is known as the tolerance). The table showing the tolerances of common Class A glassware is shown. Table 1. Tolerance for Class A Laboratory Glassware* Tolerance of Class A burets Buret volume Smallest Tolerance (mL) Graduation (mL) ((mL) 50 0.1 0.05

Tolerance of Class A transfer pipets Volume (mL) Tolerance (mL) 0.02 0.03

10 25 *Quantitative Chemical Analysis (6th ed). NY: W.H.Freeman and Company.

This is done to minimize or readily eliminate the inaccurate measurements that can be done by the instrument, and is one of the instrumentation designs fundamental aspects. With this done, an instrument that has assigned its proper measurement can now be called a standard. Also, calibration is done when the environments temperature setting is not in tune with the one from which the instrument was originally made, or formerly calibrated, since density and buoyancy factors change the slightest of a single measurement, even though the estimated significance of the Experiment 2: Calibration of Volumetric Glassware 16 | P a g e

MALAYAN COLLEGES LAGUNA

EXPERIMENT # 2 CALIBRATION OF VOLUMETRIC GLASSWARES

measurement is certain (because you are using a graduated measuring instrument). The table showing the corrections of density and buoyancy correction is shown below.

Table 2. Density of Water at Various Temperatures** Temperature (C) 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 Density (g/cm3) 0.9997026 0.9996084 0.9995004 0.9993801 0.9992744 0.9991026 0.9989460 0.9987779 0.9985986 0.9984082 0.9982071 0.9979955 0.9977735 0.9975415 0.9972995 0.9970479 0.9967867 0.9965162 0.9962365 0.9959478 Buoyancy correction mdisplay 1.00152 mdisplay 1.00152 mdisplay 1.00152 mdisplay 1.00152 mdisplay 1.00152 mdisplay 1.00152 mdisplay 1.00152 mdisplay 1.00153 mdisplay 1.00153 mdisplay 1.00153 mdisplay 1.00153 mdisplay 1.00153 mdisplay 1.00154 mdisplay 1.00154 mdisplay 1.00154 mdisplay 1.00155 mdisplay 1.00155 mdisplay 1.00155 mdisplay 1.00156 mdisplay 1.00156

30 0.9956502 mdisplay 1.00156 ** All of the densities shown for pure air-free water at a pressure of 101.325 kPa (1 atm). The buoyancy corrections assume that the air density is 1.20 x 10-3 g/cm3 and the density of the reference weight is 8.00 g/cm3. ** Analytical Chemistry and Quantitative Analysis. New Jersey: Pearson Prentice Hall. We started the experiment off by thoroughly cleaning the glassware. They are clean as it seems, but maybe the students that used it before us didnt thoroughly clean it through, so we have to do it again for assurance that the calibration will be very much accurate. The final rinsing will be Experiment 2: Calibration of Volumetric Glassware 17 | P a g e

MALAYAN COLLEGES LAGUNA

EXPERIMENT # 2 CALIBRATION OF VOLUMETRIC GLASSWARES

the distilled water, which is free of most ions present in tap water, and in effect the water will not be attracted to the sides of the glassware, creating a free flowing liquid inside. Since the distilled water that we used in the wash bottle already equilibrated to room temperature, we just read this temperature which will be used for computations later on. We started calibrating the 50-mL buret by filling the buret with distilled water. Then, by getting the mass of the Erlenmeyer flask (which will be used throughout this whole procedure) we now transfer an appropriate amount (10-ml aliquots, adding 10 to the preceding one until we have 50mL aliquots, for two trials) to the flask and weigh the flask. The difference between the last reading and the initial will be the mass of the water. After all trials and different aliquots have been done, we go to the computation step. By having the apparent mass of the water, we now compute for the corrected mass of water by the equation: ( )( ). (Eq 1)

By getting the mass, we can get the corrected/true volume by the equation: . (Eq 2)

By having the true volume, we compute for the correction value as follows: . (Eq 3) After this, we will graph the correction values vs. apparent volume. By inspection, you can see that there are negative and positive values for the correction value. This is mainly because of the difference in the mass weighed to the volume that you have estimated to the burettes significant figures (even with the help of the meniscus reader). The slope of the line will serve as the relative error of the calibration process, and if it goes by the specific tolerance, then the glassware has been properly calibrated. For the calibration of the 10-mL and 25-mL pipet, a proper container for delivery is allocated for the different measuring instruments. A 125-mL E. flask is used for the 25-mL pipet and a 50-mL beaker with parafilm for the 10-mL pipet. By getting the respective weights of the containers, for three trials we transferred distilled water to its appropriate container, and weighing it to get the difference, which will serve as the apparent mass of water. After all trials have been made, the computation process will be the same as the 50-mL buret computations, i.e. getting the correction Experiment 2: Calibration of Volumetric Glassware 18 | P a g e

MALAYAN COLLEGES LAGUNA

EXPERIMENT # 2 CALIBRATION OF VOLUMETRIC GLASSWARES

value. After this has been made, the data that we have computed and determined will be compared using statistical analysis. By getting the means of the true volumes of the different pipets, we computed for the standard deviation by the equation:

. (Eq 4)

After that, we computed for the coefficient of variation (%relative standard deviation) to see how precise our measurements are. We got a low value; therefore our precision is very high. Also, the %RSD can be used as comparison for the tolerances that is in Table 1. From what we got, we are inside the bounds of the tolerance values, therefore the values are within range. By getting the confidence interval, we computed for its actual tolerances, assuming that this is the working temperature for this glassware throughout the whole term. Lastly, we determined the %relative error, using 25-mL and 10-mL measurements as the theoretical volume. This can be compared to the tolerances, and from what we got, we are within range.

10-mL aliquots (Correction values vs. Apparent volume)


0.12 0.1 0.08 0.06 0.04 0.02 0 -0.02 9.94 0 9.96 9.98 10 10.02 0.06 Linear (Correction values) 0.1 Correction values y = 2x - 19.92

Experiment 2: Calibration of Volumetric Glassware

19 | P a g e

MALAYAN COLLEGES LAGUNA

EXPERIMENT # 2 CALIBRATION OF VOLUMETRIC GLASSWARES

20-mL aliquots (Correction values vs. Apparent volume)


0.1 0.09 0.08 0.06 0.04 0.03 0.02 0 19.85 -0.02 -0.04 Correction values Linear (Correction values) 20 20.05 -0.03 20.1 20.15 y = 0.3x - 5.97

19.9

19.95

By cleaning up everything and putting everything to what it was back before, we conclude the experiment to a finish.

IX.

SUMMARY AND CONCLUSIONS This experiment verified the concept involving the how tos of calibrating glassware, such as burettes and volumetric flasks. The glasswares were already previously calibrated during manufacture by the Quality Control (QC) department, but calibrating the instruments would be needed to ensure that no errors have been made by the QC, if ever. Even in a controlled laboratory setting, a random error can always happen. Certain factors that can affect the precision and accuracy of measurements of these glasswares are quite a few. Linear expansion would b a thing, but since the difference in the temperature in the laboratory would be very low, this can be way much disregarded. A difference in temperature, from what is said earlier, however, can affect the density of the liquid being contained or transferred. Density is the mass to volume ratio of a substance i.e. liquid, which is significant since for every degree of change, there would be a notable difference in the density. This is because the mass of

Experiment 2: Calibration of Volumetric Glassware

20 | P a g e

MALAYAN COLLEGES LAGUNA

EXPERIMENT # 2 CALIBRATION OF VOLUMETRIC GLASSWARES

the water is affected by buoyancy. With this changes, plus environmental factors, will cause a significant change between the volume obtained, and the true volume. Overall, by getting the relative difference through statistical methods and comparing it to the standard values (with the tolerance values as a guide for maximum allowable error), if it is within the bounds, then we can conclude that the glassware are calibrated to accuracy, and can be used for future purposes.

X.

POST LAB QUESTIONS 1. What is the maximum allowable error for the respective volumetric glassware that you calibrated? Based on our data, the maximum allowable error is about 0.41%, since the average tells this much difference, as to the computation o the relative error. 2. Are your errors within the tolerance volumes for the Class A glassware? What are the systematic or random errors that have occurred? Yes. If there are errors, what have possibly occurred would be: the reading of the lower meniscus, which always results to estimation; and possibly the water drops left inside the glassware due to the surface tension inside.

XI.

REFERENCES Christian, Gary D. 2004. Analytical chemistry (6th ed.). John Wiley and Sons Inc. Hage, David S. and James D. Carr. 2011. Analytical chemistry and quantitative analysis. New Jersey: Pearson Prentice Hall. Skoog, Douglas et. al. 2004. Fundamentals of Analytical Chemistry (8th ed.). Singapore: Thomson Learning.

Experiment 2: Calibration of Volumetric Glassware