Professional Documents
Culture Documents
ECA / EUSA
EG 15-14-1.2
Page 1 of 21 Rev. 0 June 1999
Instrument Calibration
(This practice is appropriate for attachment to Inquiry or Purchase Document)
SCOPE
I 1.1 This guide covers calibration standards and recommended requirements for calibration procedures of pressure and temperature instruments. Frequency of calibration is recommended, but acceptable site practices will govern. Devices such as turbine meters, positive displacement meters, coriolis meters, capacitance level devices, etc. are not covered by this guide. The fundamental accuracy of flow elements, such as orifice plates, is not covered in this guide.
I I
1.2 1.3
GENERAL
R 3.1 The purpose of calibration procedures is to optimize the accuracy of devices that measure process variables (e.g., transmitters, transducers). Calibration procedures should be (1) simple and efficient, and (2) repeatable, regardless of time, location, and personnel. Calibration results are to be documented. Primary plant calibration standards are to be used to regularly check working standards and portable transfer standards, which in turn are used to check the operating instruments in the plant.
3.2
DEFINITIONS
I 4.1 Primary Plant Standards are measurement standards that are directly traceable to national or international standards and are maintained in a protected environment in the primary standards shop. Except for recertification, this primary plant standard should not leave the standards shop. Primary Standards Shop is a facility where primary calibration standards are maintained. This shop should be clean, dust-free, humidity-controlled, and maintained at a temperature of 72F, 4F. Working Standard is a high-quality calibration standard which remains in the protected environment of the instrument shop. It should be directly compared with and proved against the primary plant standards. It should be used in the checking of portable transfer standards. Portable Transfer Standard is a portable calibration device intended for shop and field calibration of operational field transmitters and measurement equipment. These standards should have digital read-outs of 4-1/2 digits and an accuracy of at least two times better and resolution at least four times better than the field instruments on which they are used for calibration. Cardinal Calibration Points are defined as 0 percent, 50 percent, 75 percent, and 100 percent of the instrument range (except for square root extracting transmitters). Field Calibration is a calibration check and adjustment carried out with the transmitter or sensing device located in its permanent plant position. Bench Calibration is performed with the transmitter or sensing device removed from its permanent plant location and mounted in a controlled shop environment in an orientation identical to its field mounting configuration. Linearity is the difference between a straight-line transmitter output (or transducer response) at the cardinal points in the range and the actual output (response) during upscale excursion. Hysteresis is the difference between output measured at the cardinal points during an upscale excursion and the output measured at the cardinal points on the subsequent downscale return. Accuracy is the maximum deviation of the transmitter output (or transducer response) from the desired value during upscale or downscale excursion, whichever is greater. It is normally expressed as a percent of full scale, as determined at the cardinal points but may also be expressed in percent of reading or parts per million (ppm). The term limits of uncertainty is a synonym for accuracy. Optimum Calibrated Accuracy is the calibration which would give best accuracy in an instruments normal operating range. Generally, this will be in the mid 50 percent of its calibrated span. For differential pressure transmitters, Optimum Calibrated Accuracy should include final zero adjustments which may be required to compensate for high static line pressures.
I I
4.2 4.3
4.4
I I I
I I I
4.11
EG 15-14-1.2
Page 2 of 21 Rev. 0 June 1999
ECA / EUSA
4.12
4.13
Traceability is the unbroken chain of comparison of measurement apparatus emanating from the Bureau of International Physical Measurements (BIPM) in France through the various National Bureaus of Standards through industrial metrology laboratories down to operational measurement equipment. 4 Digit Meter is a digital indicator which is capable of displaying all ten digits (0 to 9) in the four least significant places but can only display a limited number of digits in the most significant place. An example is a milliammeter which, on the 20 mA DC range, cannot digitize above 19.999 mA without loss of least significant digit. The digit refers to the 1 since in this place in the display, the only digits which will appear are either a blank or 1.
CALIBRATION HIERARCHY
R 5.1 A hierarchy of primary plant standards should be maintained to compare against the working and transfer standards used for normal calibration and checkout tasks. These primary plant standards should encompass the major variables measured within the plant including analog signals. Principal parameters are pressure, temperature, and electronic, (volts, ohms, and current):
Electronic Pressure Temperature Figure 1 Figure 2 Figure 3
6.2
6.3
6.4
ENGINEERING GUIDE
ECA / EUSA
EG 15-14-1.2
Page 3 of 21 Rev. 0 June 1999
Instrument Calibration
(This practice is appropriate for attachment to Inquiry or Purchase Document)
E R
7.2 7.3
7.4
7.5
8.2
8.3
EG 15-14-1.2
Page 4 of 21 Rev. 0 June 1999
ECA / EUSA
8.4
For fast checking of operational thermoelements, an isothermal block may be employed which can easily be made in the plant (see Figure 4). Into one end is inserted the plant standard thermoelement and into the other is inserted the thermoelement under test. The block can be heated or cooled to any arbitrary temperature and then have the insulation slipped over it, the end cap insulation installed and the thermoelements inserted and their output measured for comparison purposes.
CALIBRATION FREQUENCY
R 9.1 Frequency of calibration of field devices is related to equipment, application, and experience. Calibration checks and time intervals are dictated by one or more of the following criteria: a. If the periodic zero check dictates readjustment of the zero, a complete recalibration should be scheduled. b. Substantive evidence is provided by operations which indicates that the transmitters output is suspect. c. Contractual obligations call for more frequent complete calibration checks. d. Accurate record keeping via calibration reports indicate that calibration intervals may be extended or must shortened.
FIELD-MOUNTED INSTRUMENTS R I 10.2 10.3 All instrument components should be calibrated as separate stand-alone devices. System calibrations where multiple components sequentially receive and transmit or modify a signal should be avoided. During calibration of pressure, differential pressure and pneumatic instruments, temporary connections should preferably be made using plastic or brass compression fittings. The purpose for this is: a. For ease of making up connections without leaks. To avoid damage to the instruments threaded connections during calibration (e.g., galling of stainless steel threads). All transfer standards used in calibration of field equipment should have digital readouts (4 digits minimum). Unless overriding considerations exist, zero adjustment should take place at 25 percent of range (rising) and span adjustment shall take place at 75 percent of range (rising). One exception is square-root extracting differential pressure flow transmitters where the zero adjustment should take place at 10 percent of P range and span adjustment should take place at 75 percent of P range. (NOTE: Zero adjustment refers to the process of making the transmitter output at 25% of range agree with a known value, or at 10% for square-root extracting transmitters.) Transmitter calibrations should be made only at ambient temperatures between 40F and 90F. Calibrations made outside these limits should be considered temporary and a recalibration performed when conditions permit. Permanent calibrations for transmitters in custody transfer, energy and/or material balance services should be made on the bench in the instrument shop. Where calibrations are carried out by contractor, the contractor should have his own primary plant standards and should follow the procedures defined in this standard. If the contractor does not have primary plant standards, his portable transfer standards should be checked against the plants primary standards using the procedures in this standard. b.
R R
10.4 10.5
10.6
R R
10.7 10.8
ENGINEERING GUIDE
ECA / EUSA
EG 15-14-1.2
Page 5 of 21 Rev. 0 June 1999
Instrument Calibration
(This practice is appropriate for attachment to Inquiry or Purchase Document)
TEMPERATURE TRANSMITTERS R 10.9 Accurate calibration of thermocouple amplifier transmitters and A/D converters depends on how well known the ambient temperature is. If the ambient temperature is known only to within 2F, 1C, and then the entire calibration has a margin of error of at least 2C. Therefore, calibration of such equipment should take place only in an environment where ambient temperature is stable and is measured with the desired calibration accuracy. A portable digital temperature indicator which has been checked against the primary plant temperature standards should be used for measuring ambient temperature. When calibrating resistance thermometer amplifiers and transmitters, determine the appropriate curve of each particular RTD with its associated alpha value. Use caution when assuming an alpha value.
10.10
ALARM AND TRIP SWITCHES R R R 10.11 10.12 10.13 All switches which initiate alarms or trips should be set using a rising signal for high activation and a falling signal for low activation. Prior to setting, a switch device should be exercised twice over 0 to 125 percent of its stated range. Where switch/trip devices are equipped with pointers or actuation index devices, the positioning of such a pointer or index should not replace proper calibration.
CONTROL CENTER INSTRUMENTS R 10.14 Optimum calibration of transmitters and transducers providing measurement signals as input to control center equipment requires that the transmitter be calibrated to read most accurately at or near the normal operating value. This by definition requires that accuracy of zero and 100 percent full-scale level signals be sacrificed. Therefore, a transmitter zero output should not be adjusted to conform to the control centers interpretation of what the zero output should be. However, the control centers zero reading should conform to the zero output reading of the transmitter, as measured by the transfer standard, to an accuracy of 0.016 mA or 0.1 percent F.S. The control center inputs, including all analog signals to control center equipment such as multiplexed A/D converters, distributed instrument system inputs, input module racks, etc. should be checked for calibration as follows: a. At least once annually and preferably twice annually b. R R 10.16 10.17 When disparities are detected between transmitter signals measured during periodic zero and calibration checks. All calibration checks and adjustments of control center inputs should be made using the working standard voltage/current source which has been checked against the primary standards. Control center outputs should be checked at least once annually with portable transfer standards. Disparities greater than 0.25 percent F.S. should be considered as unacceptable.
10.15
I/P TRANSDUCERS, PNEUMATIC VALVE POSITIONERS, ELECTROPNEUMATIC POSITIONERS R 10.18 High accuracy calibration and record keeping is normally not a requirement with these devices. However, a hand-held mA source of reasonable accuracy which has been checked periodically against the plant standards is a necessary part of the checkout.
I I
11.3 11.4
11.5
EG 15-14-1.2
Page 6 of 21 Rev. 0 June 1999
ECA / EUSA
I I
11.6 11.7
Vent the high- and low-pressure side of the transmitter. Close the equalizing valve and connect the pressure source to the high-pressure side of the transmitter, closing all atmospheric vents on the high-pressure side while leaving the low-pressure side vented to atmosphere. Exercise the transmitter by applying approximately 125 percent of full scale differential pressure twice (i.e., 0 percent, 125 percent, 0 percent, 125 percent, and 0 percent) to the high-pressure side. While at 125 percent of differential pressure range, check for leaks by turning off the pressure source and looking for a drop in pressure. If the pressure drops no more than 5 percent in 30 seconds, the calibration may proceed, otherwise, the leak must be found and eliminated. Pressures which should be applied during calibration are the cardinal points: 0.0 percent, 25 percent, 50 percent, 75 percent, and 100 percent of the intended calibration range of the transmitter. Final zero adjustment should take place at 25 percent of range rising, and span adjustment shall take place at 75 percent of range rising. Make initial zero adjustments at 0.0 percent of the intended calibrated differential pressure range as follows: while applying zero pressure input, adjust the zero to produce 0.0 percent, 1.0 percent of the appropriate output signal (e.g., 4.00 mA 0.16 mA). Apply 100 percent of the intended calibrated range and adjust the span to obtain 100 percent 1.0 percent of the proper full-scale output (e.g., 20.00 0.16 mA DC). Repeat the above two steps alternately until both zero and full-scale outputs are within 1.0 percent of the appropriate values. Final zero and span adjustments are 25 percent and 75 percent may now proceed. In a manner similar to the above, apply alternately 25 percent and 75 percent of the intended calibrated range. Make final zero adjustments at 25 percent of input (rising) and final span adjustments at 75 percent of input (rising). Alternate between 25 percent and 75 percent making zero and span adjustments until the transmitter output is corrected to within 0.05 percent as read by the electronic transfer standard (e.g., for 4 to 20 mA DC transmitter, 25 percent output shall be 8.00 0.008 mA DC and 75 percent output shall be 16.00 0.008 mA DC). The calibration sheet may now be filled in by applying sequentially 0 percent, 25 percent, 50 percent, 75 percent, 100 percent, 75 percent, 50 percent, 25 percent, and 0 percent of the calibrated range of the transmitter and recording the output at each value to a resolution of 0.001 mA DC. These input and output values should be recorded on the calibration sheet. Reinstall the transmitter if this calibration has taken place on the bench. Reconnect all impulse lines and plug all vents. The foregoing calibration has been accomplished with atmospheric static pressure on the low side of the capsule. After the calibration, as described, has been accomplished, the instrument zero at full line static pressure shall be checked. This can be accomplished as follows: a. At the manifold valve at the transmitter, open the high-pressure and equalizing valves. Close the low-pressure manifold valve. b. Open the primary process block valves. c. d. e. Visually check for leaks and record the static line pressure to an accuracy of 20 percent. Observe the instrument output, and record this output as Zp on the calibration sheet.
I I
11.8 11.9
11.10
11.11
I I I
11.15
I I I
I I
11.19 11.20
If the indicated zero output (Zp) as measured by the transfer standard has shifted by more than 0.1 percent relative to Za (i.e., 0.016 mA for a 4-20 mA DC output transmitter), readjust the zero to obtain an output which equals Za as recorded on the calibration sheet which was filled in during atmospheric calibration. f. Determine what control center zero reading is at this point and enter this value on the calibration sheet. Maximum allowable difference shall be 0.1 percent. If this is exceeded, the control center input circuit shall be checked. At the transmitter manifold, close the equalizing valve and open the low-pressure valve. If the transmitter is in liquid or condensible vapor service and is located below the primary element, the vents on the measurement capsule body must be cracked open to bleed off air and noncondensibles and fill the body with condensate, seal fluid, or process fluid. The transmitter is now ready for service.
11.21
ENGINEERING GUIDE
ECA / EUSA I 11.22
EG 15-14-1.2
Page 7 of 21 Rev. 0 June 1999
Instrument Calibration
(This practice is appropriate for attachment to Inquiry or Purchase Document)
Calibration procedures for square root-extracting P flow transmitters are the same as for non-square-root extracting transmitters except the cardinal points are 0 percent, 10 percent, 25 percent, 50 percent, and 100 percent of differential pressure range. Proper outputs for these points are as follows:
DIFFERENTIAL PRES. % 0 10 25 50 100 OUTPUT % 0 31.63 50.00 70.71 100.00 mA DC VALUE FOR 4-20 mA TRANSMITTER 4.000 9.061 12.000 15.314 20.000
Basic calibration adjustments of zero and span should be made at 10 percent and 100 percent of the differential pressure range. However, final full line pressure zero adjustment has a greater tolerance than with non-square-root extracting and should be 1.0 percent of output (e.g., 0.16 mA DC for a 4-20 mA transmitter). This same tolerance applies when periodic zero checks are made.
Note: For Configurable transmitters such as the Honeywell ST-3000, periodic zero checks should be made by temporarily reconfiguring the transmitter to linear in which case the criteria for non-square-root extracting applies.
The following paragraphs contain the procedure for performing periodic zero check on differential pressure transmitters. a. Using a transfer standard, read the transmitter output and refer to the calibration report for the last full calibration performed on the transmitter. Check that the operating full line pressure is within 20 percent of value used for the last full calibration. b. At the transmitter, close off the manifold valve on the low pressure side of the measuring capsule and open the equalizing valve, thus placing full equal line pressure on both sides of the transmitter. c. The transmitter output should read Za 0.1 percent (i.e., Za 0.016 mA DC). If this is the case, no further action need be taken. Proceed to paragraphs f and g. d. If the zero output is outside of these limits, the technician should again check to determine that the static line pressure is within 20 percent of the static line pressure used for the last full calibration. If the line pressure is within 20 percent, the transmitter should be replaced. e. If the static line pressure is more than 20 percent different than that used for the last complete calibration, shut off all block valves and using the vent plugs in the transmitter body vent both high and low sides to atmosphere. Refer to the calibration sheet for the value Zp and Za. The transmitter should now produce an output equal to [(2 x Za) Zp] 0.016 mA. If so, no further adjustments are required. If the indicated zero output shows a deviation greater than 0.016 mA, readjust the zero to equal (2 x Za) Zp. The transmitter is faulty and should be repaired or replaced. f. After zero check in Paragraph c, or rezeroing in Paragraphs d and e, determine what the control center zero reading is and enter this value on the calibration sheet. g. Plug all vents and close the equalizing valve and open the high and low pressure shut-off valves at the transmitter manifold. h. The transmitter is now back in service (on a temporary basis pending complete recalibration if the criteria for Za given in Paragraph c has not been met).
EG 15-14-1.2
Page 8 of 21 Rev. 0 June 1999
ECA / EUSA
CALIBRATION REPORTS
I 14.1 A documentation system should be utilized which generates calibration records for the following categories: a. b. c. d. e. I I 14.2 14.3 Transmitters. Switches. Critical temperature elements. Calibration of control center devices. Calibration checks of portable standards against the primary and working standards.
Instrument calibration sheets should be filled out for all formal calibrations and recalibrations performed by plant or contractor personnel. Following figures can be used for calibration documentation:
Pressure - Portable Transfer Standard Check Electrical Portable Transfer Standard Check Temperature Portable Transfer Standard Check Basic Transmitter Calibration Sheet D/P Transmitter Calibration Sheet (Low Pressure Side at Atmosphere) D/P Transmitter Calibration Sheet (Low Pressure Side at Line Pressure) Alarm, Switching, and Trip Instrument Calibration Sheet Critical Temperature Element Calibration Check Calibration Sheet for Control Center Instruments Figure 5 Figure 6 Figure 7 Figure 8 Figure 9 Figure 10 Figure 11 Figure 12 Figure 13
6/99
ENGINEERING GUIDE
ECA / EUSA
EG 15-14-1.2
Page 9 of 21 Rev. 0 June 1999
Instrument Calibration
(This practice is appropriate for attachment to Inquiry or Purchase Document)
Primary Plant Standard Unsaturated Voltaic Cell Null Detector and Kelvin Varley Divider Ohms Ohms MV Volts Working Standard Digital Multimeter Working Standard Decade Box
Volts
IR
MV
Volts Milliamps
pH Calibrators
MV Volts Milliamps
Ohms
EG1514012F1
EG 15-14-1.2
Page 10 of 21 Rev. 0 June 1999
ECA / EUSA
Gravity Correction
Primary Plant Standard Hydraulic Dead Weight Tester 430" WC to 10,000 psig
Gravity Correction
In-Plant Pressure and Differential Pressure Transmitter Inputs In Plant Pneumatic Analog and Control Signals
EG1514012F2
ENGINEERING GUIDE
ECA / EUSA
EG 15-14-1.2
Page 11 of 21 Rev. 0 June 1999
Instrument Calibration
(This practice is appropriate for attachment to Inquiry or Purchase Document)
or
EG1514012F3
EG 15-14-1.2
Page 12 of 21 Rev. 0 June 1999
ECA / EUSA
Test Thermoelement
EG1514012F4
ENGINEERING GUIDE
ECA / EUSA
EG 15-14-1.2
Page 13 of 21 Rev. 0 June 1999
Instrument Calibration
(This practice is appropriate for attachment to Inquiry or Purchase Document)
FIGURE 5 PRESSURE PORTABLE TRANSFER STANDARD CHECK Pressure: 4.0 to 1500 inches W.C. Working Standard Used Pneumatic Dead Weight Tester Mfr: ________________________________________ Ser. # ______________________________________ Rated Accuracy_______________________________ = AS Date of last check against Primary Standard __________________________ Transfer Standard Checked Mfr. ________________________________________ Ser. # ______________________________________ Rated Accuracy ______________________________ = AT
WORKING STANDARD TRANSFER STANDARD DEVIATION % OF SETTING = A1 % OF F.S. = A2
Disposition (If not accepted) Hold for Recalibration _________________ Return to Factory ____________________ Correction Factors applied as Follows Reading ____________________ ____________________ ____________________ ____________________ ____________________ Correction ____________________ ____________________ ____________________ ____________________ ____________________ Date _______________ Date _______________ Other ____________________ Abandon __________________
EG 15-14-1.2
Page 14 of 21 Rev. 0 June 1999
ECA / EUSA
FIGURE 6 ELECTRICAL PORTABLE TRANSFER STANDARD CHECK DIGITAL MULTIMETERS AND CALIBRATORS (Use two sheets for calibrators, one for input mode and one for output mode) Portable Transfer Standard Checked Mfr: ________________________________________ Ser. # ______________________________________ Rated Accuracy_______________________________ = AS Variables; MV Working Standards Used Mfr. ____________, ____________, ________________ Ser. # __________, _____________, _______________ Rated Accuracy _________As, ___________ As, __________ = AS WORKING STANDARDS (INPUT, OUTPUT)
MV VOLTS MA ohms
, mA
, ohms
, other
MAX DEVIATION
MV
Volts
mA
ohms
% of F.S. = As
Disposition Not Accepted; Hold for Recalibration ____________________ Return to Factory ____________________ Or Correction Factors applied as Follows Reading ____________________ ____________________ ____________________ ____________________ Correction ____________________ ____________________ ____________________ ____________________ Other ____________________ Abandon __________________
____________________ ____________________ Technician ________________________ Date ______________ Supervisor ________________________ Date ______________
ENGINEERING GUIDE
ECA / EUSA
EG 15-14-1.2
Page 15 of 21 Rev. 0 June 1999
Instrument Calibration
(This practice is appropriate for attachment to Inquiry or Purchase Document)
FIGURE 7 TEMPERATURE PORTABLE TRANSFER STANDARD CHECK Portable Transfer Standard Checked Mfr. ___________________________________ Ser. # _________________________________ Rated Accuracy ________________________ AT Type: TC , RTD , Both If RTD state alpha value ___________________ Working Standard(s) Used Mfr. ____________, ____________, ____________ Ser. # ____________, ____________, ____________ Rated Accuracy ____________ AS, ____________ AS, ____________AS Date last checked against Primary Standard _______________,_______________, _______________
WORKING STANDARD
TRANSFER STANDARD
DEVIATIONS
% OF SETTING
% OF F.S. = A2
MV
Ohms C/F
MV
Ohms C/F
MV
Ohms C/F
MV
Ohms C/F
MV
Ohms C/F
Acceptance Criteria A2 AT + AS Accepted: Disposition (If not accepted) Hold for Recalibration _______________________ Return to Factory __________________________ Correction Factors applied as Follows Reading ____________________ ____________________ ____________________ ____________________ ____________________ Yes No
Comments
Other ____________________ Abandon __________________ Correction ____________________ ____________________ ____________________ ____________________ ____________________ Date _____________ Date _____________
EG 15-14-1.2
Page 16 of 21 Rev. 0 June 1999
ECA / EUSA
FIGURE 8 BASIC TRANSMITTER CALIBRATION SHEET Tag No. _______________________________ Service # ______________________________ Xmtr Mfgr. _____________________________ Xmtr S.N. ______________________________ Calibrated Range ________________________ Calibration Units _________________________ EXISTING CALIBRATION CHECK NOTE: Cycle twice 0%/125%/0%/125%/0%, prior to starting check. Do not tap meter during test INPUT PORTABLE TRANSFER STANDARDS USED MODULE XMTR. OUTPUT Input Output INPUT Mfgr. Date Last Checked Model Ser. No. OUTPUT Mfgr. Model Ser. No. Date Last Checked Date _________________________________ Control Center _________________________ Model No. _____________________________ Instr. Spec No. _________________________
% 0 25 50 75 100 75 50 25 0
NOTE: If different calibration equipment is used here, provide mfgr. Model and serial numbers. Span shall be set at 75% rising. Zero shall be set at 25% rising. INPUT MODULE OUTPUT NOTES: Temp. of Transmitter: Temp. of Measurement Source: Temp. of Signal Monitor: Control Center Zero Reading: 5F 5F 5F
Final Calibration Check (As Left) Technician _________________________ Supervisor _________________________ Date _________ Date _________
ENGINEERING GUIDE
ECA / EUSA
EG 15-14-1.2
Page 17 of 21 Rev. 0 June 1999
Instrument Calibration
(This practice is appropriate for attachment to Inquiry or Purchase Document)
FIGURE 9 d/b TRANSMITTER CALIBRATION SHEET (LOW PRESSURE VENTED TO ATMOSPHERE) Tag No. _______________________________ Service # ______________________________ Xmtr Mfgr. _____________________________ Xmtr S.N. ______________________________ Calibrated Range ________________________ Calibration Units _________________________ EXISTING CALIBRATION CHECK NOTE: Cycle twice 0%/125%/0%/125%/0%, prior to starting check. Do not tap meter during test INPUT MODULE OUTPUT PORTABLE TRANSFER STANDARDS USED Mfgr. Model Ser. No. Mfgr. Model Ser. No. Date Last Checked Date _________________________________ Control Center _________________________ Model No. _____________________________ Instr. Spec No. _________________________
XMTR.
FINAL CALIBRATION FIELD SHOP NOTE: If different calibration equipment is used here, provide mfgr. Model and serial numbers. Span shall be set at 75% rising. Zero shall be set at 25% rising. INPUT MODULE OUTPUT NOTES:
XMTR.
Temp. of Transmitter: 5F % Input Output Temp. of Measurement Source: 5F 0 * Temp. of Signal. Monitor: 5F 25 Static Line Pressure (dp Transmitter): 20% 50 Zp = , 2 x Za Zp = 75 Control Center Zero Reading: 100 75 50 25 0 Existing Calibration (As Left) Technician ___________________________ Date _______________
Supervisor ___________________________
*This value = Za Za = Zero output at atmospheric pressure Zp = Zero output at full line static pressure
Date _______________
EG 15-14-1.2
Page 18 of 21 Rev. 0 June 1999
ECA / EUSA
FIGURE 10 d/p TRANSMITTER CALIBRATION SHEET (LOW PRESSURE SIDE AT ANTICIPATED STATIC LINE PRESSURE) Tag No._________________________________ Service # _______________________________ Xmtr Mfgr. ______________________________ Xmtr S.N._______________________________ Calibrated Range_________________________ Calibration Units__________________________ EXISTING CALIBRATION CHECK NOTE: Cycle twice 0%/125%/0%/125%/0%, prior to starting check. Do not tap meter during test XMTR. % 0 25 50 75 100 75 50 25 0 Existing Calibration Check (As Found) FINAL CALIBRATION FIELD SHOP
OUTPUT Mfgr.___________________ Date Last Checked Model__________________ __________________
Date_________________________________ Control Center _________________________ Model No. ____________________________ Instr. Spec No. ________________________
Input
Output
PRIMARY OR PORTABLE TRANSFER STANDARDS USED Mfgr.___________________ Date Last Checked Model__________________ __________________ Ser. No. ________________
NOTE: If different calibration equipment is used here, provide mfgr., model and serial numbers. Span shall be set at 75% rising. Zero shall be set at 25% rising. INPUT MODULE OUTPUT
NOTES: Temp. of Transmitter: Temp. of Measurement Source: Temp. of Signal. Monitor: Static Line Pressure (Low Pressure Reference): Control Center Zero Reading: 5F 5F 5F
Existing Calibration (As Left) Technician _________________________ Supervisor _________________________ Date _________ Date _________
ENGINEERING GUIDE
ECA / EUSA
EG 15-14-1.2
Page 19 of 21 Rev. 0 June 1999
Instrument Calibration
(This practice is appropriate for attachment to Inquiry or Purchase Document)
FIGURE 11 ALARM, SWITCHING AND TRIP INSTRUMENT CALIBRATION SHEET Tag No. __________________________ Service # _________________________ Switch Mfgr. ______________________ Model No. ________________________ Range ___________________________ Calibration Units ___________________ Type Switch: Date _________________________ Control Center ___________________ Use:
SPST SCR
Number of Setpoints:
One
Differential: Measured Input:
Model ______________
Output; How Measured _________________________ Desired Setting (Circle One) Set Point 1 _____ (Rising/Falling) Set Point 2 _____ (Rising/Falling) Set Point 3 _____ (Rising/Falling) Comments Differential _______ Differential _______ Differential _______ Set Point 1 ______/______ Set Point 2 ______/______ Set Point 3 ______/______ *Calibrated Setting Differential _______ Differential _______ Differential _______
* Enter both process value and analog signal value as applicable, e.g., 60 psig/18.5 MADC Technician ______________________ Supervisor ______________________ Date: ______ Date: ______
EG 15-14-1.2
Page 20 of 21 Rev. 0 June 1999
ECA / EUSA
FIGURE 12 CRITICAL TEMPERATURE ELEMENT CALIBRATION CHECK Tag No. ______________ Mfg. _________________ Element Type; Thermocouple J ,K ,T ,R ,S Other ______________ RTD alpha ___________ Base Resistance @ 0C ________ohms How Checked; a) Freeze Point Cell b) Temp Bath c) Isothermal Block
TABLE 1 C/F
Required Output
RTD/TC
Measured Output
RTD/TC
For Thermocouples; Ambient Temp How Checked _________________________________________ Millivolts or ohms; working or Transfer Standard used Mfr. _______________ Ser. No. _______________ Date last checked against Primary ___________ Interpret Table 1 above in C/F error at stated temperatures C/F ____ ____ ____ ____ Accepted State Criteria Technician _____________________________ Date: ____________ Supervisor _____________________________ Date: ____________ Error _____ _____ _____ _____ Yes No
ENGINEERING GUIDE
ECA / EUSA
EG 15-14-1.2
Page 21 of 21 Rev. 0 June 1999
Instrument Calibration
(This practice is appropriate for attachment to Inquiry or Purchase Document)
Quantity of I or O
Input Range/Parameter, 4-20/MADC / How Checked: Working Standard/Portable Standard (Circle One) Standard Used: Mfg. Model No. Ser. No. Date Last Checked
MVDC /
Reading of Input or Output 0% Required I/O Value Circuit Tag No. Slot No. or Other Identifier Measured or Indicated Value ____________ 25% ____________ 50% ____________ 75% ____________ 100% ____________